178
SVAN 958A User Manual
results of the measurement from the 3
rd
profile of the 2
nd
channel if the LOGGER MODE was set to ON
(path: MENU / INPUT / LOGGER SETUP / LOGGER MODE:ON); and if any position in CHAN. 2
PROF. 3 (path: MENU / INPUT / LOGGER SETUP / CHANNEL 2 / CHAN. 2 PROF. 3) sub-list was
selected, up to four words are written in the given sequence:
<result1> - PEAK result in the case of SLM if the first position was marked, else no value is written;
<result2> - MAX result in the case of SLM if the second position was marked, else no value is written;
<result3> - MIN result in the case of SLM if the third position was marked, else no value is written;
<result4> - RMS result in the case of SLM if the fourth position was marked, else no value is written;
results of the measurement from the 3
rd
profile of the 3
rd
channel if the LOGGER MODE was set to ON
(path: MENU / INPUT / LOGGER SETUP / LOGGER MODE:ON) and if any position in CHAN. 3
PROF. 3 (path: MENU / INPUT / LOGGER SETUP / CHANNEL 3 / CHAN. 3 PROF. 3) sub-list was
selected, up to four words are written in the given sequence:
<result1> - PEAK result in the case of SLM if the first position was marked, else no value is written;
<result2> - MAX result in the case of SLM if the second position was marked, else no value is written;
<result3> - MIN result in the case of SLM if the third position was marked, else no value is written;
<result4> - RMS result in the case of SLM if the fourth position was marked, else no value is written;
results of the measurement from the 3
rd
profile of the 4
th
channel if the LOGGER MODE was set to ON
(path: MENU / INPUT / LOGGER SETUP / LOGGER MODE:ON) and if any position in CHAN. 4
PROF. 3 (path: MENU / INPUT / LOGGER SETUP / CHANNEL 4 / CHAN. 4 PROF. 3) sub-list was
selected, up to four words are written in the given sequence:
<result1> - PEAK result in the case of SLM if the first position was marked, else no value is written;
<result2> - MAX result in the case of SLM if the second position was marked, else no value is written;
<result3> - MIN result in the case of SLM if the third position was marked, else no value is written;
<result4> - RMS result in the case of SLM if the fourth position was marked, else no value is written;
VECTOR measurement result if in the LOGGER MODE (path: MENU / INPUT / LOGGER SETUP /
LOGGER MODE:ON) and VECTOR (path: MENU / INPUT / LOGGER SETUP / VECTOR:ON) are set
to ON and VECTOR measurement was enabled; one word is written.
RPM measurement result if the LOGGER MODE (path: MENU / INPUT / LOGGER SETUP /
LOGGER MODE:ON) and RPM (path: MENU / INPUT / LOGGER SETUP / AUXILIARY/ RPM:[
])
are set to ON and RPM measurement was enabled; two word are written.
results of 1/1 OCTAVE analysis from the 1
st
channel if 1/1 OCTAVE analysis was selected as the
measurement function and in the LOGGER (path: MENU / INPUT / 1/1 OCTAVE SETUP /
CHANNEL 1: ENABLED[
]; LOGGER:RMS) other then None value was selected; the sequence of
words is written:
<flags> <Octave[1]> <Octave[2]> ... <Octave[NOct+NOctTot]>
where:
flags = 1 - the overload detected, 0 - the overload not detected
Octave[i] - the result of 1/1 OCTAVE analysis (*10 dB); i = 1..NOct+NOctTot (1..18)
results of 1/1 OCTAVE analysis from the 2
nd
second channel if 1/1 OCTAVE analysis was selected as
the measurement function and in the LOGGER (path: MENU / INPUT / 1/1 OCTAVE SETUP /
CHANNEL 2 / ENABLED[
]; LOGGER:RMS) other then None value was selected; the sequence of
words is written:
<flags> <Octave[1]> <Octave[2]> ... <Octave[NOct+NOctTot]>
where:
flags = 1 - the overload detected, 0 - the overload not detected

Table of Contents