5.4.5 Initializing Service Settings
(1) Press [Int. Settings-2] key, then the confirmation screen with “Initialize Entire Memory?” message will
appear. When selecting this key, following service setting items will be initialized in order.
Service Settings
Analysis Parameters
Confirmation Mode
Reagent Vial Settings
Gain Correction Coefficient
(2) Select the appropriate key.
[Set]: To start service setting initializing. When completing, the “Initial” screen will return.
[Cancel]: To cancel service setting initializing, and return to the “Initial” screen.
5.4.6 Data Check
(1) Select “Verify Control Tbl” key, then the confirmation screen with “Initialize Entire Memory?” message will
appear.
(2) Select the appropriate key.
[Set]: To perform memory check in order of following screen. The check result will be indicated “OK”
or “NG”. The number on the screen “8333” indicates the checksum value of Flash Memory.
[Cancel]: To cancel memory check, and return to the “Initial” screen.
(3) When [Set] is pressed, memory check will be performed in the following order.
Flash Memory
Motor position adjustment value
(Detector Block adjustment, Temperature control, Pressure adjustment)
Manufacturer settings (Service settings, Analysis Parameters, Reagent Bottle)
Customer settings
Standard Curve
Each sample data in stored data
Each file data in QC data
8
Important: Note that any NG file is automatically initialized.
Stored Data
Standard Data
User Set
Maker Set
Position
Verification Control Table in progress
Return
Initialize
IPReady
Replace Rack? YES!
Sysmex
Quality Control
OK
OK
OK
OK
OK
OK
8333
Figure 5-4-3: Data Check Screen
5.4.7 Contents stored in BBRAM
(1) Stored data
(2) QC data
(3) Flag for error message “Turned Off During Operation”.
Syringe operation cycle counts
Test counts
(4) Group selection number
Rinse Wait time (Clean Sample Probe)
Catcher transfer test cycle counts
Catcher transfer test error cycle counts
Storing dummy range
CA-500 Series S/M 5-7 Revised December 2001