6.7 ANALYSIS ERRORS
6.7.1 Analysis Error
Monitor Timing: When measuring/analyzing
Monitor Method:
Coagulation
Error code Status Action
ERROR0001
(Temp Error)
Temperature error occurred when
sample was being analyzed.
Refer to 6.2 Temperature Errors.
ERROR0002
(Slight Coagulation)
Detected coagulation curve shows
extremely small change.
Reanalyze the sample to make
comprehensive judgment, taking sample,
reagent, etc. into account.
ERROR0004
(Analysis Time
Over)
Coagulation did not finish within
preset detection time or within Auto
Extension Mode detection time.
Reanalyze the sample to make
comprehensive judgment, taking sample,
reagent, etc. into account. Or conduct
another test with prolonged detection time.
ERROR0008
(Coag. Curve Error)
1. Curve shows a drop in some area.
2. Curve shows too steep in some
areas.
3. As buffer was cool, clotting time
was extended by the influence of
air bubbles. (Fbg)
ERROR0016
(Turbidity Level
Over)
Turbidity was too high to analyze. Reanalyze the sample by pre-dilution.
ERROR0032
(No coagulation)
Clotting could not be detected. Reanalyze the sample after checking
samples and reagents (storage status blood
collecting method, etc.) for any errors, and
make comprehensive judgment.
ERROR0064
(Measurement
Error)
Due to any problem before photo
detection, such as insufficient sample
plasma, instrument error, insufficient
reagent, Mechanical Stop Switch
operation, analysis could not be done.
Reanalyze the sample after preparing
required amount of plasma.
ERROR0100
(Range Over)
The detection time at 50% point is
exceeding the preset time, i.e.
Reanalyze the sample after checking
samples and reagents (storage status,
ERROR0200
(Fbg Range Over)
coagulation is drastic. blood collecting method, etc.) for any errors,
and make comprehensive judgment.
Chromogenic/Immunoassy
Error code Status Action
ERROR0401
(Trans Light Low)
Transmitted Light was too low. Check the reagent, lamp, etc. and reanalyze.
ERROR0402
(Trans Light High)
Transmitted Light was too high. Check the reagent, lamp, etc. and reanalyze.
ERROR0404
(No Linearity)
Reaction Curve had no linearity. Reanalyze the sample.
ERROR0408
(Reaction Curve Error)
Reaction Curve changed to the
opposite direction.
Check the setting, reagent and reanalyze.
ERROR0416
(Range Over)
Antigen became range over (with
Immunoassy only).
Reanalyze the sample by re-dilution, etc.
ERROR0528
(No Polynomial
Adjustment)
Polynomial adjustment could not
be performed (when using
VlinIntegral mode only).
Check the setting, reagent and reanalyze.