6.9.9 HC ACK Time Out
6.9.10 HC STX Time Out
6.9.11 HC ETX Time Out
Monitor Timing: During data transmission
Monitor Method: Monitors RS-232C protocol and time out
CA-500’s Action: Warning
(1) Probable Cause
1. Failure of connection cable
2. Failure of host computer
3. Failure of PCB No. 6362
(2) Confirmation by the operator
1. Check the connection with host computer
2. Check the host computer.
(3) Confirmation by Service Engineer
1. Replace the PCB No. 6362.
6.9.12 Instruction Not Found in HC
Monitor Timing: During data transmission
Monitor Method: Monitoring RS-232C protocol and time out
CA-500’s Action: Warning
(1) Probable Cause
1. No registration of Work List.
(2) Confirmation by the operator
1. Register the Work List in the host computer correctly.
(3) Confirmation by Service Engineer
The same as above (2).
6.9.13 Serial Illegal Error
Monitor Timing: During data transmission
Monitor Method: Monitoring RS-232C protocol and time out
CA-500’s Action: Warning
(1) Probable Cause
1. Receive Data failure caused by noise, etc.
2. Problem is occurred on the program of the host computer.
(2) Confirmation by the operator
1. Register the Work List in the host computer correctly.
(3) Confirmation by Service Engineer
The same as above (2).
CA-500 Series S/M 6-33 Revised December 2001 8