6.9.4 HC Communication Error
6.9.5 HC CTS Time Out
Monitor Timing: During data transmission
Monitor Method: Monitors RS-232C protocol and time out
CA-500’s Action: Warning
(1) Probable Cause
1. Failure of connection cable
2. Failure of host computer
3. Failure of PCB No. 6362
(2) Confirmation by the operator
1. Check the connection with host computer.
2. Check the host computer.
(3) Confirmation by Service Engineer
1. Replace the PCB No. 6362.
6.9.6 HC ACK Code Error
6.9.7 HC Transmission Count Error
6.9.8 HC Reception Count Error
Monitor Timing: During data transmission
Monitor Method: Monitors RS-232C protocol and time out
CA-500’s Action: Warning
(1) Probable Cause
1. Failure of connection cable
2. Failure of host computer
3. Failure of PCB No. 6362
(2) Confirmation by The operator
1. Check the connection with host computer.
2. Check the host computer.
(3) Confirmation by Service Engineer
1. Replace the PCB No. 6362.
CA-500 Series S/M 6-32 Revised December 2001 8