EasyManua.ls Logo

SYSMEX CA-500 - 4.3.5 Position Adjustment Procedure of Pipette

SYSMEX CA-500
259 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CA-500 Series S/M 4-16 Revised December 2001
4.3.5 Position Adjustment Procedure of Pipette (“[4]reag 1”) (“[5]reag 9”) (“[6]reag 10”)
Continuous Z position adjustment is possible on software version 00-14 or later. Refer to 4.3.6-2 11
(1) XYZ Adjustment
1) Set the sample cup holder to the reagent rack No. 1. Place 60 µL of distilled water into the 4 mL
sample cup.
Figure 4-3-8: Sample Cup with 60 µL Distilled Water
2) Verify that the “[4]reag 1” is shown on the left center of the screen. If not, change to “[4]reag 1” by
using [Pos. -] and [Pos. +] keys.
3) Move the pipette to the setting position by using [XY TEST] key.
4) Use [n], [p], [m], and [Æ] keys to adjust the pipette tip to become the center of the sample cup.
Figure 4-3-9: Pipette Tip Position
5) Press [Z TEST] key to descent the pipette. The pipette detects the 60 µL of distilled water and stops.
6) Use [n], [p], [m], and [Æ] keys to adjust the pipette tip to become the center of the sample cup.
7) Press [Z ADJUST] key to ascend and descend the pipette. Verify the pipette is placed at the center
of the sample cup.
8) Press [XY ORG] key to return the pipette (head) to the home position.
9) Adjust “[5]reag 9” and “[6]reag 10” in the same way.
CAUTION: 1) Descent the pipette lower limit to the lowest for Z axis adjustment value. Adjust
after executing Section 4.3.4: Position Adjustment Procedure of Pipette, (3)
Pipette Lower Limit Adjustment when it stops before detecting liquid surface.
Set “650” pulse for the pipette lower limit of the reagent rack.
8

Table of Contents

Related product manuals