EasyManua.ls Logo

SYSMEX CA-500 - 4.3.6 Position Adjustment Procedure of Pipette

SYSMEX CA-500
259 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CA-500 Series S/M 4-17 Revised December 2001
4.3.6 Position Adjustment Procedure of Pipette (“[7]reag_rinse”) (“[8] buffer”)
Continuous Z position adjustment is possible on software version 00-14 or later. Refer to 4.3.6-2 11
(1) XY Adjustment
1) Set the sample cup holder to the CA CLEAN I Holder. Place 60 µL of distilled water into the 4 mL
sample cup.
Figure 4-3-10: Sample Cup with 60 µL Distilled Water
2) Verify that the “[7]reag_rinse” is shown on the left center of the screen. . If not, change to
“[7]reag_rinse” by using [Pos. -] and [Pos. +] keys.
3) Move the pipette to the setting position by using [XY TEST] key.
4) Use [n], [p], [m], and [Æ] keys to adjust the pipette tip to become the center of the sample cup.
Figure 4-3-11: Pipette Tip Position
5) Press [Z TEST] key to descent the pipette. The pipette detects the 60 µL of distilled water and stops.
6) Use [n], [p], [m], and [Æ] keys to adjust the pipette tip to become the center of the sample cup.
7) Press [Z TEST] key again to ascend and descend the pipette. Verify that the pipette is placed at the
center of the sample cup.
8) Press [XY ORG] key to return the pipette (head) to the home position.
9) Adjust “[8]buffer” in the same way
8

Table of Contents

Related product manuals