EasyManua.ls Logo

SYSMEX CA-500 - How to Operate

SYSMEX CA-500
259 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1.7 HOW TO OPERATE
Check
Instrument
• Check the rinse solution (distilled water) to make sure that the analyzer is ready
for operation.
Power ON
• Turn the instrument power ON 30 minutes before the analysis.
Analysis
Preparation
• Dissolve reagents 30 minutes before the analysis.
Set the reagents to the reagent holder.
Execute the quality control to check the instrument condition.
Start Analysis
• Set the sample to the sampler.
Enter sample orders to be analyzed.
Press the [Start] key to start analysis.
Result Output
• The analysis results will be displayed on LCD and printed out on printer paper.
End Analysis
• Remove the reagents and the sample.
Shutdown
• Execute the daily maintenance for the next analysis.
Power OFF • Rinse the pipette and turn the instrument power OFF.
CA-500 Series S/M 1-15 Revised December 2001 8

Table of Contents