5.2 SERVICE SETTING
Use to set the default parameter when initializing.
From the Main menu, press [Special Operate] Æ [Service] Æ [Service Setting] keys.
Next
Option
Ready
Replace Rack? YES!
Chrom.
DFbg
Immuno.
Wavelength
D-Dimer Init.
APTT Warm
ON
ON
OFF
575 nm
for DD PLUS
180 Sec.
Service Setting
Sysmex
Return
IP
Next
Figure 5-2-1: Service Setting Screen
[n] or [p]
Press to select the item to be set. Selected item is indicated in reversed.
(1) Chrom., Immuno., TTO thru PFDP
Select to set the selected parameter’s analysis protocol automatically when performing initializing.
ON: To set.
OFF: Not to set.
(2) DFbg (Derived Fbg)
Set the 4th calculation parameters of PT standard curve as:
ON: Valid
OFF: Invalid
(3) APTT Warm
When selecting this item, the numeric keys will appear.
The APTT incubation time can be set using numeric keys in 30 sec increments.
If the APTT incubation time is set, the default setting of APTT incubation time is automatically set
when performing initializing.
Numeric keys: To input value
Enter: To set the input value*. (*The input vale is verified.)
[Next Option]
Pressing this key alternates “ON” or “OFF”.
[Return]
Use this key to quit the settings, and then the following keys will appear.
[Cancel]: Cancels the renewed setting and returns to the Service Menu screen.
[FIX]: Changes to the renewed setting and returns to the Service Menu screen.
(Setting data will be stored in the memory.)
[Continue]: Returns to the Service Setting screen and allows continuous operation.
CA-500 Series S/M 5-2 Revised December 2001 8