6.11.4 Enter Work List
Monitor Timing: At time of analysis start
Monitor Method: Software
CA-500’s Action: No analysis
(1) Probable Cause
1. The Work List is not registered.
2. Operation error
(2) Confirmation by the operator
1. Register the Work List correctly.
(3) Confirmation by Service Engineer
The same as above (2).
6.11.5 Check Reagent Volume 1
Monitor Timing: At time of analysis start
Monitor Method: Monitors preset data on EEPROM and actual accumulated amount of volume
CA-500’s Action: Warning
(1) Probable Cause
1. When reagent volume monitor in valid, the theoretical calculation volume of reagent exceeds the used
reagent volume.
(2) Confirmation by the operator
1. Replenish reagent and input new data of reagent volume.
(3) Confirmation by Service Engineer
The same as above (2).
6.11.6 Insufficient Reaction Tube
Monitor Timing: At time of analysis start
Monitor Method: Software
CA-500’s Action: Does not analyze.
(1) Probable Cause
1. The number of test registered in Work List exceeds that of the remaining reaction tubes (a maximum
of 60 tests).
(2) Confirmation by the operator
1. Reduce the number of test registered in Work List.
2. Replenish the reaction tube.
(3) Confirmation by Service Engineer
The same as above (2).
CA-500 Series S/M 6-38 Revised December 2001 8