EasyManua.ls Logo

SYSMEX CA-500 - Reagent Temp; Tube Transfer Test

SYSMEX CA-500
259 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CA-500 Series S/M 5-12 Revised February 2004
19 TB200408
5.5.4 Reagent Temp. 19
Verify the reagent pipette incubation operation.
(1) Press [Special Operate] Æ [Service] Æ [Test Menu] Æ [Reagent Temp.] keys to start, then the following
screen will be displayed.
3
Quit
Enter
6
9
Reagent Temp.
Set
C
4
1
0
7
5
2
8
Volume
Ready
Replace Rack? YES!
Sysmex
IP
100 uL
TUBE 1
Figure 5-5-5: Reagent Temp. Screen
(2) To verify Reagent Incubation operation
1) Pour 50 µL of distilled water into a reaction tube. Set a thermometer into the water. Then put it into the
incubation well #6. Verify that the distilled water temperature in the reaction tube is 37±1°C.
2) Put the sample cup containing distilled water in reagent holder #1.
3) Set the dispensing volume using numeric keys at 100 µL (default value).
4) Press [Enter] key to set, eventually, the input volume to 100ul.
5) Remove the thermometer from reaction tube and press [Set] key to perform the heating operation.
6) After the reagent nozzle is dispensing 100 µL of water in incubation well #6 from reagent holder #1,
the reaction tube is moved to detection well. Confirm that the temperature of water at detection well is
within 37±1°C. It is recommended that the temperature of water at detection well is adjusted within
37±0.3°C
7) If the temperature value is not within the range, return to the Temperature adjustment procedure
section 4.4.3 input the new S-value as follows:
When the measured value is higher than the range 37±0.3°C use "+" key to add 5 to the current
Nozzle S-value.
When the measured value is smaller than the range 37±0.3°C, use "-" key to subtract 5 to from the
current Nozzle S-value.
After the S-value change, perform steps 1) – 6) again.
8) Press [Quit] key to finish the test. Then the screen will return to the Test Menu display.
5.5.5 Tube Transfer Test
Verify the reaction tube transfer operation.
(1) Press [Special Operate] Æ [Service] Æ [Test Menu] Æ [Tube Trans Test] keys to start, then the following
screen will be displayed.
Tube Trans Test
Tube No. = 1
Detect No. = 1
Warm No. = 1
Ready
Replace Rack? YES!
Sysmex
IP
ReturnSet
Figure 5-5-6: Tube Transfer Test Screen

Table of Contents

Related product manuals