EasyManua.ls Logo

SYSMEX CA-500 - 5.3 Adjust

SYSMEX CA-500
259 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.3 ADJUST
Use to adjust the hardware positions, sensitivities, etc.
Refer to Section 4 Adjustment for the adjustment procedures of each item.
From the Main menu, press [Special Operate] Æ [Service] Æ [Adjust] keys.
Return
Adjust
IPReady
Replace Rack? YES!
Sysmex
Mecha Position
Detector Gain
Temp. Control
Adjust. of Pump
Trans. Light Calib.
Adjust Data Print
Figure 5-3-1: Adjust Menu Screen
5.3.1 Printing Adjustment Data
From Adjust menu, press [Adjust Data Print] key.
The adjustment setting values are printed to the built-in printer.
06/17/1999 11:22
Position PULSE
blood1 X 1839
Y 2265
Z 1246
blood10 X 4061
Y 2265
Z 1246
blood_stat X 1485
Y 2265
Z 1246
06/17/1999 11:22
DETECT
OFFSET GAIN
1 HIGH 128 128
LOW 139 128
PT 139 128
2 HIGH 114 159
LOW 134 232
PT 134 235
warm3 X 1469
Y80
Z-356
warm4 X 605
Y944
Z-356
comp X 722
Y512
Z-369
shake X 137
Y512
Z-29
dust X 154
Y203
Z-376
06/17/1999 11:22
TEMP
COOLER 128
Main 128
Sub 128
DETECT 128
Main 128
Sub 128
PIPETTE 128
Main 128
Sub 128
ROOM 128
Main 128
Sub 128
Figure 5-3-2: Example of Setting Data Print Out
CA-500 Series S/M 5-4 Revised December 2001 8

Table of Contents

Related product manuals