EasyManua.ls Logo

SYSMEX CA-500 - Page 115

SYSMEX CA-500
259 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
The changeable parameters are as follows:
Parameter Description Key to be Used Changeable Range
Vial Name Name of Reagent Vial Full Keyboard 9 characters maximum
(Alphabets and Numbers)
Std Normal Volume (mL) Numeric Keys 0.01 thru 99.99 (by 0.01 mL)
Dead Dead Volume (mL) Numeric Keys 0.00 thru 99.99 (by 0.01 mL)
Area [mm
2
] Section Area of Vial (mm
2
) Numeric Keys 0.000 thru 999.999 (by 0.001 mm
2
)
Offset Number of Offset Pulse
from Adjustment Position
to Dead Volume
Numeric Keys -999 thru 999 (by 1 pulse)
Sysmex
Ready
Replace Rack? YES!
IP
Cancel
Vial Name
Dead
Cup
4.00 mL
0.10 mL
105.683
0
Reag. Vial Settings
Std
Area [mm ]
Offset
Adjust Z pulse down
Set
2
Figure 5-7-3: Z Pulse Down Adjustment Screen
[Set] Key
Use to execute Descend Pulse Adjustment screen.
[Cancel] Key
Cancels the Z Pulse Down Adjustment and returns to the Reagent Vial Settings screen.
<Z Pulse Down Adjustment>
(1) Fill the dead volume of the reagent in the reagent vial and place at the reagent holder No. 9.
(2) Press [Set] key to execute Z Pulse Down Adjustment.
(3) The probe moves to the reagent holder No. 9 and detect a liquid surface for 5 times.
(4) The difference between the average of the descend pulse and the value of the reagent holder No. 9
position adjustment will be calculated as the pulse correction value.
CA-500 Series S/M 5-20 Revised December 2001 8

Table of Contents

Related product manuals