(5) TTO
Sample Aspiration 20 µL
Dilution volume 12 Buffer 30 µL
First Reagent 180 sec
Reagent volume 8 TTO 125 µL
Rinse 11 125 µL
Detection Activity-TTO
Coag. detection point 50%
Sensitivity High Sensitivity
Max. detection time 190 sec
(6) NT
Sample Aspiration 20 µL
Dilution volume 12 Buffer 40 µL
First Reagent 180 sec
Reagent volume 10 NT 125 µL
Rinse 11 125 µL
Detection Activity-NT
Coag. detection point 50%
Sensitivity High Sensitivity
Max. detection time 190 sec
(7) AT-III
Sample Aspiration 10 µL
Dilution volume 12 Buffer 120 µL
Sample Aspiration 10 µL
Dilution volume 12 Buffer 10 µL
First Reagent 30 sec
Reagent volume 4 AT3.T 100 µL
Rinse 11 100 µL
Second Reagent 90 sec
Reagent volume 6 AT3.S 100 µL
Rinse 11 100 µL
Detection Chromogenic
Coag. start point 11 sec
Coag. end point 40 sec
Sensitivity Low Sensitivity
NOTE: There is a case that (4) TT - (7) AT-III are not set. It depends on the Service Settings.
CA-500 Series S/M 5-28 Revised December 2001 8