(8) Reagent Storage Function
1) Reagent Storage Method: Stored in the reagent rack
2) Reagent holder capacity: Maximum 10 kinds of reagent bottles can be set in the reagent
rack. Buffer and rinsing solution can also be set.
3) Usable container: Dade Bering’s old DADE for 5 mL, a sample cup, can be set, and
also GW5, a new type of reagent bottles only for
CA-550/CA-560, can be set.
(The outer diameter of the reagent bottle should be less than
22.5 mm, and also, if the inner diameter is too small, use the
bottle adapter.)
4) Reagent Cooling Function: Four reagents can be cooled. (CA-530, CA-540)
5) Temperature Control Accuracy: within 15 ± 2°C (Room temperature: 15°C - 30°C)
6) Waiting time for setting temp.: within 30 minutes
7) Reagent Mixing Function: None
Type of reagent Number of
accommodation
Cooling function Container which can be
accommodated
PT 1 Yes
APTT 2 No
Fbg 1 Yes DADE 5 mL Reagent Bottle
TTO 1 No (with outer diameter of
NT 1 No 22.3 mm)
TT 1 Yes
AT-III 2 Yes (only one)
Buffer 1 No
CA CLEAN I 1 No
(9) Detection Function
1) Principle of Measurement
[Coagulation Method]: Scattered Light Measuring Method (wave length 660 nm)
[Chromogenic Method]: Calorimetric method (wave length 405 nm)
[Immunoassay]: Colorimetric method (wave length 575 nm)
2) Number of Detectors
[Coagulation Method]: 4 detection wells (High and Low sensitivity automatic switching)
[Chromogenic Method]: 1 detection well
[Immunoassay]: 1 detection well
3) No. of Incubation wells: 6 wells
4) Temperature Control Accuracy: within 37.0 ± 0.5°C (ambient temperature: 15°C-30°C)
within 37.0 ± 1.0°C for Incubation Well
5) Waiting time for setting temp.: within 30 minutes
(10) Pipette Rinsing Function
1) External Container: Rinse Bottle and Waste Bottle
2) Rinse Cup: Probe Rinse Cup (Outside) and Probe Rinse Cup (Inside)
3) Rinsing Function: Pipette is rinsed by controlling the syringe and the solenoid valve, with
applying the pressure on the Rinse Bottle.
4) Waste Draining: The waste is aspirated by controlling the solenoid valve, with applying
vacuum on the waste bottle.
CA-500 Series S/M 1-10 Revised December 2001 8