Quadra 4 Operator Manual
Tomtec
109
"BAD DATA Errors" occur when the Quadra is unable to properly communicate with the
reader. Typically this is caused when the BCR jumper cable is either not connected to
bar code reader ports or a connected cable has the incorrect wiring. The Bar Code Error
Handling group is enabled whenever a bar code logging is required. The operator has
the option of ignoring an error, recording the error without intervention, or be prompted
to enter the proper bar codes on an error.
Set as Default
The current logging options and runtime database usage settings are saved as default
settings for new programs.
Reset to Default
The current logging options and runtime database usage settings are reset to default
settings for new programs.
5.9.2 Runtime Database Usage Options
When this feature is selected, the plate and fixture heights are ignored, and raw stage
heights are used. The default is to use the plate and fixture database heights, which
requires plates and fixtures present at runtime. You might want to check this if you are
moving a program to another system where plates and fixtures may not be in the
database.