EasyManua.ls Logo

TomTec QUADRA 4 - Setting Zero Volume Adjustment for Pipettor Heads

TomTec QUADRA 4
217 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Quadra 4 Operator Manual
Tomtec
170
5.0 Setting the Zero Volume Adjustment for Air
Displacement Pipettor Heads
SETTING THE ZERO VOLUME ADJUSTMENT FOR 450uL AIR TIP SHUCK
PIPETTING HEAD AND DMV AIR TIP SHUCK PIPETTING HEAD
For this procedure, a Zero Volume Setting Adjustment Gauge is required:
450uL AIR TIP SHUCK PIPETTING HEAD 026929 0.156” (3.062mm)
DMV AIR TIP SHUCK PIPETTING HEAD 026928 0.110” (2.794mm)
(1) Under “System Parameters” select “Pump Zero Volume Alignment Utility”.

Table of Contents