Quadra 4 Operator Manual
Tomtec
20
Compressor & Pneumatics
Pneumatic Panel Description
The front panel provides manual access to the pneumatic system. The pressure gauge
may be connected to several ports for readout.
Compressor Pressure: 30psi for 60 Hz operation.
25psi for 50Hz operation.
NOTE: Due to normal compressor pulsing, do not leave the pressure gauge connected
to the compressor for long time periods. This affects the life of the pressure gauge.
Blowout Pressure: This is set by the regulator on the right hand side of the
Quadra 4. It will range from 0psi to 5psi. A more accurate
pressure reading is available on the front display.
Elevator Pressure:
Regulatory Accessory (API) Pressure:
At system pressure (not a user adjustment).
This pressure is set by the R2 pressure regulator.
It must be less than 10psi or it will rupture the bladder in the various nests.
Switches: Elevator Override - Actuates the stacker elevators.
Compressor Bypass:
Shuts off the internal air compressor and connects system to external air
supply. External air supply must not exceed 35psi.