Quadra 4 Operator Manual
Tomtec
192
2.1 Calibration Curve Procedure - label each Weighing Vessel with the
sample number - Consisting of the volume to be calibrated and the pipetting number (1,
2, 3) i.e., 5-1, 5-2, 5-3, … for each of the 20 SET VOLUMES listed below required for
creating the calibration curve.
SET VOLUME 1
ST
PIPETTING 2
ND
PIPETTING 3
RD
PIPETTING
5µL 5-1 5-2 5-3
8 µL 8-1 8-2 8-3
10µL 10-1 10-2 10-3
12µL 12-1 12-2 12-3
15µL 15-1 15-2 15-3
20µL 20-1 20-2 20-3
25µL 25-1 25-2 10-3
30µL 30-1 30-2 30-3
40µL 40-1 40-2 40-3
50µL 50-1 50-2 50-3
60µL 60-1 60-2 60-3
70µL 70-1 70-2 70-3
90µL 90-1 90-2 90-3
100µL 100-1 100-2 100-3
110µL 110-1 110-2 110-3
125µL 125-1 125-2 125-3
150µL 150-1 150-2 150-3
200µL 200-1 200-2 200-3
300µL 300-1 300-2 300-3
450µL 450-1 450-2 450-3
Using the 4 Place Balance – Record tare weight of each Weighing Vessel.