The data is entered into the software with a fluid class name i.e. water, methanol,
acetonitrile, DMSO, etc. The fluid class name is not case sensitive. It may be entered
in upper or lower case letters. The date of the calibration is also entered.
The Quadra 4
®
programs are created using the fluid class name i.e. water. Any time in
the future this program is called, it will use that fluid class calibration. The actual data
will be the default value of the last data entry for that fluid class i.e. water (6-16-06).
Other fluid classes, or other dated curves, may be selected at the time the program is
executed. The default value will be the last value entered for that fluid class, unless
another selection is made at the time of running the program.
1.0 Equipment Required
Four place Calibrated Precision Balance
ISO Grade 3 distilled or de-ionized water.
Weighing Vessels – (60) Used to Create a calibration curve
(30) Used to Verify the calibration curve
Cat. No. 320-50 96 Well Tip Loading Jig
(Supplied with the QUADRA 4
)
Cat. No. 196-130-00 Polypropylene Reagent Reservoir with 3 baffles
(Supplied with the QUADRA 4
)
Cat. No. 196-205 QUADRA Pipetting Tips
(Supplied with the QUADRA 4
)
2.0 Weighing Vessels
Weighing Vessels may be purchased commercially or user made from light weight
aluminum foil. The lid of a Microtiter Plate or a 96 well Flat Bottom Microtiter Plate may
also be used.