Quadra 4 Operator Manual
Tomtec
195
A popup window “ENTER NEW PUMP CALIBRATION TABLE DATA” will appear.
Fill in the values for “EXPECTED VOLUME” and “ACTUAL VOLUME” from
Step 3 Calibration Summary” spreadsheet. Select the “ADD” button.
Continue filling in the “EXPECTED VOLUME” and “ACTUAL VOLUME” boxes for all
20 SET VOLUME results from “Step 3 Calibration Summary” spreadsheet. Select
the “OK” button.
Checking the “SHOW REFERENCE CURVE” box will graphically display the new
calibration curve.
7.0 “Calibration Template – Q4” Spreadsheet
Open the CALIBRATION TEMPLATE – Q4” Spreadsheet, making sure the
ENABLE MACROS” button is selected.
On the “Step 1 Define µL Volume to Calibration” spreadsheet select the RED
“Populate Calibration Vol. Example Data” button.
Save the spreadsheet by selecting “SAVE AS” and naming the file by the LIQUID
CLASS, and today’s date i.e “WATER (09-22-16)” XX is the today’s date.
Open the “Step 2 Calibration Data” and at the top of the spreadsheet fill in the
following information:
TODAY’S DATE
TIME
TEMP º C (To nearest 0.2ºC)
HUMIDITY % (To nearest 10%)
BAROMETRIC PRESSURE (in. Hg)
HEAD SERIAL NUMBER
Fill in the tare weight of each Calibration weighing Vessel in the area marked
“TARE WT.”.