10.3 File Management – Curve Database
The pump curve database, QCURVEDB.DAT, is UNIQUE to the Quadra 4 its assigned
to. Although this file CANNOT be shared, it can be backed up in either another folder or
on a server. Note that when backing up to a server, a separate folder must be created
for each Quadra 4. If this is not done, the saved curve database will be useless.
If multiple Quadra 4's run the same set of programs, the Curve Name may be different
between the Quadras.
A work around for this is to choose one curve and copy its name, ascii date, and binary
date to the CurveDB files in the other Quadras. You should use an ASCII text editor,
Note Pad, to make these changes.