8. SETTING
Config. Copy.
You can copy the setting parameters of one CV system to another CV system.
Connect the USB memory to the USB connector on condition that this screen is displayed.
The buttons are validated.
Import configurations
Capture the setting parameters from the USB memory.
Connect the USB memory that saves the setting parameters of CV system, and this button
is valid.
Click this button. The window showing "Configuration files have been updated. Please
restart CV software to apply new configurations." is displayed.
Click the [OK] button, and the CV software is finished. Start the CV software again.
Export configurations
Write the setting parameters to the USB memory.
Connect the USB memory, and this button is valid.
Click this button, and the window showing "Just a moment, please." is displayed. Then,
when the window showing "Configuration files have been successfully exported" is dis-
played, the procedure is finished.
When the setting file is output on condition that two or more user accounts are
set, the setting of the account that has executed the output is written.
The following items are not copied by "Import configurations" and "Export con-
figurations":
Date setting, Time setting, Next data number, Chart selection, IP address, Serial
configuration, etc.

Table of Contents

Other manuals for Topcon Compu-Vision CV-5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Topcon Compu-Vision CV-5000 and is the answer not in the manual?

Topcon Compu-Vision CV-5000 Specifications

General IconGeneral
BrandTopcon
ModelCompu-Vision CV-5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals