11. SPECIFICATIONS AND PERFORMANCE
SYSTEM CLASSIFICATION
Types of protection against electric shocks: This instrument is classified as
Class I equipment.
Class I equipment does not depend only on basic insulation for protection against electric
shocks, but also provides a means of connection to a protective earth system of facilities so
that metal parts that come into contact do not become conductive while the basic insulation is
in failure.
Degree of protection against electric shocks: Type B applied part
Type B applied part is the applied part complying with the specified requirements of the Stan-
dard IEC 60601-1 to provide protection against electric shock, particularly regarding allowable
LEAKAGE CURRENT.
The forehead rest of this instruments is "Type B applied part".
Degree of protection against harmful ingress of water: IPx0
CV-5000 has no protection against ingress of water. (The degree of protection against harmful
ingress of water defined in IEC60529 is IPx0.)
Classification according to the method(s) of sterilization or disinfection rec-
ommended by the manufacturer: not applicable.
CV-5000 has no part to be sterilized or be disinfected.
Classification according to the degree of safety of application in the pres-
ence of a flammable anesthetic mixture with air or with oxygen or nitrous
oxide: Equipment not suitable for use in the presence of a flammable anes-
thetic mixture with air or with oxygen or nitrous oxide.
CV-5000 should be used in environments where no flammable anesthetics and/or flammable
gases are present.
Classification according to the mode of operation: Continuous operation.
Continuous operation is the operation under normal load for an unlimited period, without the
specified limits of temperature being exceeded.