221
11. SPECIFICATIONS AND PERFORMANCE
ELECTROMAGNETIC COMPATIBILITY
This product conforms to the EMC standard (IEC60601-1-2 Ed.3.0: 2007).
a) MEDICAL ELECTRICAL EQUIPMENT requires special precautions regarding EMC and needs
to be installed and operated according to the EMC information provided in the ACCOMPANY-
ING DOCUMENTS.
b) Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
c) The use of ACCESSORIES, transducers and cables other than those specified, with the
exception of transducers and cables sold by the manufacturer of the EQUIPMENT or SYSTEM
as replacement parts for internal components, may result in increased EMISSIONS or
decreased IMMUNITY of the EQUIPMENT or SYSTEM.
d) The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other equip-
ment. If adjacent or stacked use is necessary, the EQUIPMENT or SYSTEM should be
observed to verify normal operation in the configuration in which it will be used.
e) The use of the ACCESSORY, transducer or cable with EQUIPMENT and SYSTEMS other than
those specified may result in increased EMISSION or decreased IMMUNITY of the EQUIP-
MENT or SYSTEM.
Item Article code
Cable
Shielded
Ferrite Core Length (m)
CV power supply unit with exter-
nal personal computer
43224 9202 — — —
Measuring head connection cable 43221 5790 Yes Yes 0.6
CV power supply relay cable 43221 5780 Yes Yes 8.0
Power cord 44804 7018 No No 3.0
DVI cable 42428 5001 Yes Yes 10.0
Communication cable 41812 0002 Yes No 5.0
Guidance and manufacturer's declaration - electromagnetic emissions
The CV-5000 is intended for use in the electromagnetic environment specified below.
The customer or the user of the CV-5000 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The CV-5000 uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby elec-
tronic equipment.
RF emissions
CISPR 11
Class B
The CV-5000 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply net-
work that supplies buildings used for domestic pur-
poses.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Complies