11. SPECIFICATIONS AND PERFORMANCE
Guidance and manufacturer's declaration - electromagnetic immunity
The CV-5000 is intended for use in the electromagnetic environment specified below.
The customer or the user of the CV-5000 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment
- guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, con-
crete or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV
line(s) to line(s)
± 2 kV
line(s) to earth
± 1 kV
line(s) to line(s)
± 2 kV
line(s) to earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
Voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
t
(>95% dip in U
t
)
for 0.5 cycle
40% U
t
(60% dip in U
t
)
for 5 cycles
70% U
t
(30% dip in U
t
)
for 25 cycles
<5% U
t
(>95% dip in U
t
)
for 5 sec
<5% U
t
(>95% dip in U
t
)
for 0.5 cycle
40% U
t
(60% dip in U
t
)
for 5 cycles
70% U
t
(30% dip in U
t
)
for 25 cycles
<5% U
t
(>95% dip in U
t
)
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the CV-5000 requires contin-
ued operation during power
mains interruptions, it is recom-
mended that the CV-5000 be
powered from an uninterruptible
power supply or battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic
fields should be at levels char-
acteristic of a typical location in
a typical commercial or hospital
environment.
NOTE U
t
is the a.c. mains voltage prior to application of the test level.

Table of Contents

Other manuals for Topcon Compu-Vision CV-5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Topcon Compu-Vision CV-5000 and is the answer not in the manual?

Topcon Compu-Vision CV-5000 Specifications

General IconGeneral
BrandTopcon
ModelCompu-Vision CV-5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals