68
5. OPERATION FOR TESTS
5.1.3 ASTIGMATISM TEST
Search for the approximate astigmatic correction value of the patient.
1 Prepare for the test.
Before beginning this test, carry out the following operation:
Set the cylinder axis "180" and cylinder power "0.00". Perform fogging in the plus direc-
tion and then shift the power in the minus direction. Adjust the spherical power to the
lowest so that the visual acuity may be 0.5 to 0.6 (20/40 to 20/30, 6/18 to 6/9).
2 Select the astigmatism test chart.
To perform the astigmatism test, select the astigmatism test chart icon on the chart page.
In the near-point test, select the proper target from the near-point card targets manually.
Example of the test chart icons on chart page
The following items are automatically changed as shown below.
Far-point
chart
Far-point test chart icon Near-point
chart
Near-point test chart icon
ACP-8(A) NC-3(E)
Item Description
Test eye
When any other item except "binocular" is the last test object,
the status of last test is succeeded. When "binocular" are the
last test object, the right eye of the patient is selected.
Auxiliary lens Not set.
Mous operation
mode
The mode to operate the cylinder axis is set.
Mouse Navigation

Table of Contents

Other manuals for Topcon Compu-Vision CV-5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Topcon Compu-Vision CV-5000 and is the answer not in the manual?

Topcon Compu-Vision CV-5000 Specifications

General IconGeneral
BrandTopcon
ModelCompu-Vision CV-5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals