112
5. OPERATION FOR TESTS
7 Record the result for the "Recovery" point.
Click the [Recovery (-)] button to record the result.
As the "Blur" point, the value to be recorded is the value obtained by adding 3.00D to the
value, which is displayed in "ADD", with the reversed sign.
5.4.4 POSITIVE RELATIVE ACCOMMODATION MEASURE-
MENT
Place the near-point visual acuity target that meets an visual acuity of approx. 0.7 (20/30, 6/9) at
40cm in front of the patient's eyes. Have the patient view this target. As adding the concave
spherical lenses, check to what degree the patient can see the target clearly. By conducting this
test, you are measuring the positive accommodative power for the patient.
1 Select "[N] Positive Relative Accommodation [#20]" from "Examination list".
To perform this test, specify the test name from "Examination list".
Click the [Examination list] button to display "Examination list".
Click the [Far/Near] button to display the near-point test list.
Click the [Bino] tab and select "[N] Positive Relative Accommodation [#20]" from the list.