8.
Verify that the that the
dot and ring projections are centred to the primary
collimator beam
(if not refer to paragraph 11.2.11.4)
9.
Then in Exam parameters window select: ID: Centring Ceph; Format: Height
24; Resolution High; Params1: Static Acquisition Centered; kV; 60; mA: 2; time:
300ms
10.
Take an X-ray emission and verify that the x-ray field is centered in the sensor
area as shown in the following figure
NOTE to better view the sensor area,zoom in by clicking 2 times on the + icon in
the program bar