Service Manual Installation
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
120
8.
Verify that the that the
dot and ring projections are centred to the primary
collimator beam
(if not refer to paragraph 11.2.11.4)
9.
Then in Exam parameters window select: ID: Centring Ceph; Format: Height
24; Resolution High; Params1: Static Acquisition Centered; kV; 60; mA: 2; time:
300ms
10.
Take an X-ray emission and verify that the x-ray field is centered in the sensor
area as shown in the following figure
NOTE to better view the sensor area,zoom in by clicking 2 times on the + icon in
the program bar