Service Manual Troubleshooting
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
201
9.2.10.3
E1403: Software watchdog
This message is displayed if the software did not periodically reset the Ethernet
watchdog timer.
It may be related to a wrong machine configuration (e.g. the ceph option has
been disabled).
Refer to paragraph 8.4 to check machine configuration.
Refer to Error E1402 (see paragraph 9.2.10.2).
Technical Service additional information required
:
-
Software logs
-
3D Sensor logs folder (see paragraph 11.2.1).