EasyManua.ls Logo

Acteon X-MIND prime 3D ceph - Page 222

Acteon X-MIND prime 3D ceph
376 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Manual Periodic maintenance
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
220
34.
Open the "PhD_C_Test.exe" service program and make an exposure at 60kV
2.2mA without objects in the X-ray field.
35.
Verify that there are no defect lines or inhomogeneous bands on the
panoramic image:
36.
Select a ceph30x24LL, High resolution, position the ceph head holder in latero
lateral position, place the nasion in field and take an exposure at 60kV3.2mA
without objects in the X-ray field.
37.
Verify that there are no defect lines or inhomogeneous horizontal bands on
the image.
38.
Select a ceph30x24LL, Normal resolution, position the ceph head holder in
latero lateral position, place the nasion in field and take an exposure at
60kV3.2mA without objects in the X-ray field.
39.
Verify that there are no defect lines or inhomogeneous horizontal bands on
image.
40.
Open AIS software and open a test patient
41.
Perform the following 3D acquisitions at 60kV 2.2mA using the relevant chin
support:

Table of Contents

Related product manuals