Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-29
SCPI Commands
:DISPlay[:WINDow][:STATe]
:DISPlay[:WINDow][:STATe]
This command controls whether the instrument screen is updated or not.
AT *RST, this value is ON.
Syntax :DISPlay[ :WINDow] [ :STATe] OFF | ON | 0 | 1
Parameter
Query response 0 | 1 <newline><^END>
Remarks The :DISP OFF (or 0) command is not available when the 4155C/4156C screen
displays the following page:
• SYSTEM: FILER
• SYSTEM: MISCELLANEOUS
• SYSTEM: CONFIGURATION
• SYSTEM: SELF-CALIBRATION/DIAGNOSTICS
• SYSTEM: PRINT/PLOT SETUP
• SYSTEM: COLOR SETUP
• KNOB SWEEP
To enter the following command, the 4155C/4156C screen must be set to the update
enable state.
• :PAGE:KSWeep command group
• :PAGE:SCONtrol:KSWeep[:STARt] command
• :DIAGnostic:TEST[:EXECute] test_no command (test_no: 201 to 413)
Example OUTPUT @Hp4155;":DISP ON"
OUTPUT @Hp4155;":DISP?"
ENTER @Hp4155;A
Parameter Type Explanation
OFF or 0 boolean screen is not updated
ON or 1 boolean screen is updated (update enable state)