EasyManuals Logo

Agilent Technologies 4155C Command Reference

Agilent Technologies 4155C
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #111 background imageLoading...
Page #111 background image
Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-29
SCPI Commands
:DISPlay[:WINDow][:STATe]
:DISPlay[:WINDow][:STATe]
This command controls whether the instrument screen is updated or not.
AT *RST, this value is ON.
Syntax :DISPlay[ :WINDow] [ :STATe] OFF | ON | 0 | 1
Parameter
Query response 0 | 1 <newline><^END>
Remarks The :DISP OFF (or 0) command is not available when the 4155C/4156C screen
displays the following page:
• SYSTEM: FILER
• SYSTEM: MISCELLANEOUS
• SYSTEM: CONFIGURATION
• SYSTEM: SELF-CALIBRATION/DIAGNOSTICS
• SYSTEM: PRINT/PLOT SETUP
• SYSTEM: COLOR SETUP
• KNOB SWEEP
To enter the following command, the 4155C/4156C screen must be set to the update
enable state.
• :PAGE:KSWeep command group
• :PAGE:SCONtrol:KSWeep[:STARt] command
• :DIAGnostic:TEST[:EXECute] test_no command (test_no: 201 to 413)
Example OUTPUT @Hp4155;":DISP ON"
OUTPUT @Hp4155;":DISP?"
ENTER @Hp4155;A
Parameter Type Explanation
OFF or 0 boolean screen is not updated
ON or 1 boolean screen is updated (update enable state)

Table of Contents

Other manuals for Agilent Technologies 4155C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155C and is the answer not in the manual?

Agilent Technologies 4155C Specifications

General IconGeneral
BrandAgilent Technologies
Model4155C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals