Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-239
SCPI Commands
:PAGE:MEASure[:SWEep]:VARD:PCOMpliance:STATe
:PAGE:MEASure[:SWEep]:VARD:PCOMpliance:STA
Te
This command sets the power compliance of VAR1' to disable.
If SMU for VAR1' unit is set to VPULse or IPULse mode, STATe is set to OFF.
You cannot set power compliance for VAR1'.
If the unit type of VAR1' is VSU, this parameter is ignored.
You use this command only if there is an SMU whose function (FCTN) is VAR1'.
If power compliance value is set by PCOMpliance pcompliance, the query returns
1.
At *RST, this value is OFF.
Syntax :PAGE:MEASure [ :SWEep] :VARD:PCOMpliance:STATe OFF | 0
Parameter
Query response 0 | 1 <newline><^END>
Example OUTPUT @Hp4155;":PAGE:MEAS:VARD:PCOM:STATE OFF"
OUTPUT @Hp4155;":PAGE:MEAS:VARD:PCOM:STATE?"
ENTER @Hp4155;A
Parameter Type Explanation
OFF or 0 boolean power compliance is disabled