EasyManuals Logo

Agilent Technologies 4155C Command Reference

Agilent Technologies 4155C
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #237 background imageLoading...
Page #237 background image
Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-155
SCPI Commands
:PAGE:KSWeep:VAR1:POINts
:PAGE:KSWeep:VAR1:POINts
This command sets the number of steps for VAR1 of knob sweep measurement.
At *RST, this value is 101.
Syntax :PAGE:KSWeep:VAR1:POINts points | MINimum | MAXimum
Parameter
Query response points <newline><^END>
points is NR1 response data type.
Example OUTPUT @Hp4155;":PAGE:KSW:VAR1:POIN 101"
OUTPUT @Hp4155;":PAGE:KSW:VAR1:POIN MAX"
OUTPUT @Hp4155;":PAGE:KSW:VAR1:POIN?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:KSW:VAR1:POIN?MAX"
ENTER @Hp4155;A
Parameter Type Explanation
points numeric 1 to 1001