5-270 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:STRess:SETup:HTIMe
:PAGE:STRess:SETup:HTIMe
This command sets the hold time to wait between the non-synchronous (non-stress)
channel output and the synchronous channel (stress) output.
At *RST, this value is 0.
Syntax :PAGE:STRess:SETup:HTIMe hold_time | MINimum | MAXimum
Parameter
Query response hold_time <newline><^END>
hold_time is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:STR:SET:HTIM 0.5"
OUTPUT @Hp4155;":PAGE:STR:SET:HTIM MAX"
OUTPUT @Hp4155;":PAGE:STR:SET:HTIM?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:STR:SET:HTIM? MAX"
ENTER Hp4155;A
:PAGE:STRess:SETup[:MENU]
This command changes the present display page to STRESS: STRESS SETUP.
This command does not have query form.
Syntax :PAGE:STRess:SETup [ :MENU]
Example OUTPUT @Hp4155;":PAGE:STR:SET"
:PAGE:STRess:SETup:MODE
This command selects the stress mode.
If no PGU is set to VPULse mode, stress mode must be set to DURation.
At *RST, this value is DURation.
Syntax :PAGE:STRess:SETup:MODE DURation | PCOunt
Parameter Type Explanation
hold_time numeric 0 to 655.35 (s)

Table of Contents

Other manuals for Agilent Technologies 4155C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155C and is the answer not in the manual?

Agilent Technologies 4155C Specifications

General IconGeneral
BrandAgilent Technologies
Model4155C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals