5-154 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:KSWeep:VAR1:MODE
Example OUTPUT @Hp4155;":PAGE:KSW:VAR1:EXT 0.1"
OUTPUT @Hp4155;":PAGE:KSW:VAR1:EXT?"
ENTER @Hp4155;A
:PAGE:KSWeep:VAR1:MODE
This command sets the SWEEP MODE of VAR1 for knob sweep measurement.
At *RST, this value is SINGle.
Syntax :PAGE:KSWeep:VAR1:MODE SINGle | DOUBle
Parameter
Query response SING | DOUB <newline><^END>
Example OUTPUT @Hp4155;":PAGE:KSW:VAR1:MODE SING"
OUTPUT @Hp4155;":PAGE:KSW:VAR1:MODE?"
ENTER @Hp4155;A$
Parameter Type Explanation
SINGle character single stair sweep
DOUBle character double stair sweep