Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-179
SCPI Commands
:PAGE:MEASure:PGUSetup:PULSe:PGU<n>:BASE
:PAGE:MEASure:PGUSetup:PULSe:PGU<n>:BASE
This command sets the BASE VALUE of PGU<n>.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
You use this command only if the specified PGU is set to VPULse mode (not V) by
the :PAGE:CHAN:PGU<n>:MODE command.
At *RST, this value is 0.
Syntax :PAGE:MEASure:PGUSetup:PULSe:PGU<n>:BASE base | MINimum |
MAXimum
Parameter
Query response base <newline><^END>
base is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:BASE 1"
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:BASE MIN"
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:BASE?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:BASE? MIN"
ENTER @Hp4155;A
Parameter Type Explanation
base numeric -40 to 40 V