EasyManuals Logo

Agilent Technologies 4155C Command Reference

Agilent Technologies 4155C
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #130 background imageLoading...
Page #130 background image
5-48 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:HCOPy:ITEM[:WINDow]:TEXT2:STATe
:HCOPy:ITEM[:WINDow]:TEXT2:STATe
This command sets whether the GRAPH AXIS TEXT (names, units, and scale of
the graph axis) should be plotted or printed when the :HCOPy:IMMediate command
or :HCOPy:DATA? query is sent.
At *RST, the value of this parameter is ON.
This command does not comply with SCPI.
Syntax :HCOPy:ITEM [ :WINDow] :TEXT2:STATe OFF | ON | 0 | 1
Parameter
This command has meaning only when :HCOPy:DEVice:LANGuage is set to
HP-GL. If set to PCL, GRAPH AXIS TEXT is always printed or plotted.
Query response 0 | 1 <newline><^END>
Example OUTPUT @Hp4155;":HCOP:ITEM:TEXT2:STAT ON"
OUTPUT @Hp4155;":HCOP:ITEM:TEXT2:STAT?"
ENTER @Hp4155;A
:HCOPy:ITEM[:WINDow]:TEXT3:STATe
This command sets whether the measurement setup data should be plotted or printed
when the :HCOPy:IMMediate command or :HCOPy:DATA? query is sent to output
graphics results.
At *RST, the value of this parameter is OFF.
This command does not comply with SCPI.
Syntax :HCOPy:ITEM [ :WINDow] :TEXT3:STATe OFF | ON | 0 | 1
Parameter Type Explanation
OFF or 0 boolean GRAPH AXIS TEXT is not plotted or printed.
ON or 1 boolean GRAPH AXIS TEXT is plotted or printed.

Table of Contents

Other manuals for Agilent Technologies 4155C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155C and is the answer not in the manual?

Agilent Technologies 4155C Specifications

General IconGeneral
BrandAgilent Technologies
Model4155C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals