EasyManuals Logo

Agilent Technologies 4155C Command Reference

Agilent Technologies 4155C
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #250 background imageLoading...
Page #250 background image
5-168 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure:MSETup:ITIMe:SHORt[:APERture]
:PAGE:MEASure:MSETup:ITIMe:SHORt[:APERture
]
This command sets the SHORT integration time in terms of the APERTURE, which
means time.
The parameter has units of seconds.
At *RST, this value is 640 ms.
Syntax :PAGE:MEASure:MSETup:ITIMe:SHORt [ :APERture] aperture | MINimum |
MAXimum
Parameter
Query response aperture <newline><^END>
aperture is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:SHORT 1E-3"
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:SHORT MIN"
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:SHORT?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:SHORT? MIN"
ENTER @Hp4155;A
Parameter Type Explanation
aperture numeric 8E-5 to 1.92E-3 (s)