5-152 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:KSWeep:DISPlay:REGion:X
:PAGE:KSWeep:DISPlay:REGion:X
This command sets the display region of X axis for knob sweep.
At *RST, this value is POSitive.
Syntax :PAGE:KSWeep:DISPlay:REGion:X POSitive | NEGative | BIPolar
Parameter
Query response POS | NEG | BIP <newline><^END>
Example OUTPUT @Hp4155;":PAGE:KSW:DISP:REG:X POS"
OUTPUT @Hp4155;":PAGE:KSW:DISP:REG:X?"
ENTER @Hp4155;A$
:PAGE:KSWeep:DISPlay:REGion:Y
This command sets the display region of Y axis for knob sweep.
At *RST, this value is POSitive.
Syntax :PAGE:KSWeep:DISPlay:REGion:Y POSitive | NEGative | BIPolar
Parameter
Query response POS | NEG | BIP <newline><^END>
Example OUTPUT @Hp4155;":PAGE:KSW:DISP:REG:Y POS"
OUTPUT @Hp4155;":PAGE:KSW:DISP:REG:Y?"
ENTER @Hp4155;A$
Parameter Type Explanation
POSitive character display positive X region
NEGative character display negative X region
BIPolar character display positive and negative X region
Parameter Type Explanation
POSitive character display positive Y region
NEGative character display negative Y region
BIPolar character display positive and negative Y region