Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-191
SCPI Commands
:PAGE:MEASure:QSCV:CONStant:SMU<n>[:SOURce]
:PAGE:MEASure:QSCV:CONStant:SMU<n>[:SOURce]
This command sets the SOURCE value of the constant source SMU<n> for the
Quasi-static CV measurement. This command also has a query form.
Where <n> is required to specify the SMU number. Valid SMU numbers are SMU1
through SMU6, depending on the configuration.
Use this command only if the function of the specified SMU is CONStant and the
mode is not COMMon.
At *RST, this value is set to 0.
Syntax :PAGE:MEASure:QSCV:CONStant:SMU<n>[ :SOURce] source | MINimum |
MAXimum
:PAGE:MEASure:QSCV:CONStant:SMU<n>[ :SOURce]?
Parameter
Query response source <newline><^END>
source is an NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:QSCV:CONS:SMU1 10"
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:CONS:SMU1?"
ENTER @Hp4155;A
Parameter Type Explanation
source numeric –200 to 200 V or –1 to 1 A. The range of this value
depends on the type of SMU being used.