5-228 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure[:SWEep]:VAR1:PCOMpliance:STATe
:PAGE:MEASure[:SWEep]:VAR1:PCOMpliance:STA
Te
This command sets the power compliance of VAR1 to disable.
If SMU for VAR1 unit is set to VPULse or IPULse mode, STATe is set to OFF. You
cannot set power compliance for VAR1.
If the unit type of VAR1 is VSU, STATe is ignored.
If power compliance value is set by PCOMpliance pcompliance, the query returns
1.
At *RST, this value is OFF.
Syntax :PAGE:MEASure [ :SWEep] :VAR1:PCOMpliance:STATe OFF | 0
Parameter
Query response 0 | 1 <newline><^END>
Example OUTPUT @Hp4155;":PAGE:MEAS:VAR1:PCOM:STATE OFF"
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:PCOM:STATE?"
ENTER @Hp4155;A
Parameter Type Explanation
OFF or 0 boolean power compliance is disabled