Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-83
SCPI Commands
:PAGE:CHANnels[:CDEFinition]:ALL:DISable
:PAGE:CHANnels[:CDEFinition]:ALL:DISable
This command deletes the settings of all units (SMU,VSU,VMU,PGU,GNDU).
This command does not have query form.
Syntax :PAGE:CHANnels [ :CDEFinition] :ALL:DISable
Example OUTPUT @Hp4155;":PAGE:CHAN:ALL:DIS"
:PAGE:CHANnels[:CDEFinition]:COMMent
This command sets the USER DEFINED COMMENT for the measurement group.
At *RST, a comment is not defined.
Syntax :PAGE:CHANnels [ :CDEFinition] :COMMent comment
Parameter
Query response comment <newline><^END>
comment is string response data.
Example OUTPUT @Hp4155;":PAGE:CHAN:COMM'V-Icurve'"
OUTPUT @Hp4155;":PAGE:CHAN:COMM?"
ENTER @Hp4155;A$
Parameter Type Explanation
comment string String of up to 58 characters.