5-298 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:STATus:QUEStionable:NTRansition
:STATus:QUEStionable:NTRansition
This command sets the negative transition filter of the Questionable Status
"CONDITION" register.
Syntax :STATus:QUEStionable:NTRansition register
Parameter
Query response register <newline><^END>
register is a decimal integer value, which is the sum of the binary-weighted values,
in NR1 response data format.
Semantics If a bit in the negative transition filter is set to 1, then a 1 to 0 transition in the
corresponding bit of the Questionable Status "CONDITION" register causes a 1 to
be written in the corresponding "EVENT" register.
Example OUTPUT @Hp4155;":STAT:QUES:NTR 16384"
OUTPUT @Hp4155;":STAT:QUES:NTR #H4000"
OUTPUT @Hp4155;":STAT:QUES:NTR #Q40000"
OUTPUT @Hp4155;":STAT:QUES:NTR #B100000000000000"
OUTPUT @Hp4155;":STAT:QUES:NTR?"
ENTER @Hp4155;A
Parameter Type Explanation
register numeric or
non-decim
al numeric
decimal integer, hexadecimal, octal, or binary value
that is the sum of the binary-weighted values for the
desired bits.