5-256 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:STRess[:CDEFinition]:SELector<n>:STRess
:PAGE:STRess[:CDEFinition]:SELector<n>:STRess
This command selects the relay connection mode of Agilent 16440A SMU/PGU
selector n for stress state.
<n> is required to specify the SMU/PGU selector number. Valid SMU/PGU selector
numbers are SELector1 through SELector4.
At *RST, this value is OPEN.
Syntax :PAGE:STRess [ :CDEFinition] :SELector<n>:STRess SMU | PGU | OPEN |
POPen
Parameter
Query response SMU | PGU | OPEN | POP <newline><^END>
Example OUTPUT @Hp4155;":PAGE:STR:SEL1:STR SMU"
OUTPUT @Hp4155;":PAGE:STR:SEL1:STR?"
ENTER @Hp4155;A$
:PAGE:STRess[:CDEFinition]:SMU<n>:DISable
This command deletes the stress settings of SMU<n>.
<n> is required to specify SMU number. Valid SMU numbers are SMU1 through
SMU6, depending on the configuration.
This command does not have query form.
Syntax :PAGE:STRess [ :CDEFinition] :SMU<n>:DISable
Example OUTPUT @Hp4155;":PAGE:STR:SMU1:DIS"
Parameter Type Explanation
SMU character connect to SMU port
PGU character connect to PGU port
OPEN character disconnect all
POPen character disconnect all, but disconnect PGU port with
semiconductor switch, not normal switch. This
parameter is valid only for selector 1 and 3.