5-244 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:SCONtrol:STANdby
:PAGE:SCONtrol:STANdby
This command controls the output of standby channels.
This command can be executed only when the present state is IDLE or STANDBY.
But the query can be executed for any state.
Executing this command is about the same as pressing Standby front-panel key.
At *RST, this value is OFF.
Syntax :PAGE:SCONtrol:STANdby OFF | ON | 0 | 1
Parameter
Query response 0 | 1 <newline><^END>
Example OUTPUT @Hp4155;":PAGE:SCON:STAN ON"
OUTPUT @Hp4155;":PAGE:SCON:STAN?"
ENTER @Hp4155;A
:PAGE:SCONtrol:STATe?
This command returns the present state of the 4155C/4156C.
This command has query form only.
Syntax :PAGE:SCONtrol:STATe?
Query response IDLE | STAN | MEAS | STR <newline><^END>
Example OUTPUT @Hp4155;":PAGE:SCON:STAT?"
ENTER @Hp4155;A$
Parameter Type Explanation
OFF or 0 boolean change to IDLE state
ON or 1 boolean change to STANDBY state
Parameter Type Explanation
IDLE character IDLE state
STAN character STANdby state
MEAS character MEASurement state
STR character STRess state

Table of Contents

Other manuals for Agilent Technologies 4155C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4155C and is the answer not in the manual?

Agilent Technologies 4155C Specifications

General IconGeneral
BrandAgilent Technologies
Model4155C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals