5-224 Agilent 4155C/4156C SCPI Command Reference, Edition 1
SCPI Commands
:PAGE:MEASure[:SWEep]:PULSe:PERiod
:PAGE:MEASure[:SWEep]:PULSe:PERiod
This command sets the pulse PERIOD of SMU.
You use this command only if an SMU is set to VPULse or IPULse mode.
Syntax :PAGE:MEASure [ :SWEep] :PULSe:PERiod period | Minimum | MAXimum
Parameter
Query response period <newline><^END>
period is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:PULS:PER 0.1"
OUTPUT @Hp4155;":PAGE:MEAS:PULS:PER MAX"
OUTPUT @Hp4155;":PAGE:MEAS:PULS:PER?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:PULS:PER? MAX"
ENTER @Hp4155;A
:PAGE:MEASure[:SWEep]:PULSe:WIDTh
This command sets the pulse WIDTH of SMU.
You use this command only if an SMU is set to VPULse or IPULse mode.
Syntax :PAGE:MEASure [ :SWEep] :PULSe:WIDTh width | MINimum | MAXimum
Parameter
Query response width <newline><^END>
width is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:MEAS:PULS:WIDTH 0.1"
Parameter Type Explanation
period numeric 5E-3 to 1 (s). Resolution: 100 ms.
Parameter Type Explanation
width numeric 5E-4 to 1E-1 (s). Resolution: 100 ms.