Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-259
SCPI Commands
:PAGE:STRess[:CDEFinition]:SMU<n>:NAME
:PAGE:STRess[:CDEFinition]:SMU<n>:NAME
This command sets the NAME of SMU<n> for stress.
<n> is required to specify SMU number. Valid SMU numbers are SMU1 through
SMU6, depending on the configuration.
At *RST, this value is:
Syntax :PAGE:STRess [ :CDEFinition] :SMU<n>:NAME name
Parameter
Query response name <newline><^END>
name is string response data, but does not contain double quote characters at the
beginning and end of the string.
Example OUTPUT @Hp4155;":PAGE:STR:SMU1:NAME 'VD'"
OUTPUT @Hp4155;":PAGE:STR:SMU1:NAME?"
ENTER @Hp4155;A$
SMU NAME
SMU1 through SMU4 V<n>
SMU5 and SMU6 not defined
Parameter Type Explanation
name string String of up to 6 alphanumeric characters.
1st character must be alphabet.