Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-277
SCPI Commands
:PAGE:STRess:SETup:PULSe:PGU<n>:WIDTh
Parameter
Query response trailing_time <newline><^END>
trailing_time is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:TRA 0.001"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:TRA MAX"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:TRA?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:TRA? MAX"
ENTER @Hp4155;A
:PAGE:STRess:SETup:PULSe:PGU<n>:WIDTh
This command sets the pulse WIDTH of PGU<n> for stress.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
The mode of the specified PGU must be VPULse.
At *RST, this value is 5 ms.
Syntax :PAGE:STRess:SETup:PULSe:PGU<n>:WIDTh width | MINimum | MAXimum
Parameter
Query response width <newline><^END>
width is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:WIDTH 0.005"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:WIDTH MIN"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:WIDTH?"
ENTER @Hp4155;A
Parameter Type Explanation
trailing_time numeric 100 ns to 10 ms
Parameter Type Explanation
width numeric 1E-6 to 9.99 s