EasyManua.ls Logo

Agilent Technologies 4155C

Agilent Technologies 4155C
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 4155C/4156C SCPI Command Reference, Edition 1 5-265
SCPI Commands
:PAGE:STRess:SETup:CONStant:PGU<n>[:SOURce]
Example OUTPUT @Hp4155;":PAGE:STR:SET:ACC RES"
OUTPUT @Hp4155;":PAGE:STR:SET:ACC?"
ENTER @Hp4155;A
:PAGE:STRess:SETup:CONStant:PGU<n>[:SOURce]
This command sets the constant SOURCE value of PGU<n> for stress.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
You use this command only if the specified PGU is set to V mode (not VPULSE) by
the :PAGE:STR:PGU<n>:MODE command.
At *RST, this value is 0.
Syntax :PAGE:STRess:SETup:CONStant:PGU<n> [ :SOURce] source | MINimum |
MAXimum
Parameter
Query response source <newline><^END>
source is NR3 response data type.
Example OUTPUT @Hp4155;":PAGE:STR:SET:CONS:PGU1 10"
OUTPUT @Hp4155;":PAGE:STR:SET:CONS:PGU1 MAX"
OUTPUT @Hp4155;":PAGE:STR:SET:CONS:PGU1?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:STR:SET:CONS:PGU1? MAX"
ENTER @Hp4155;A
Parameter Type Explanation
source numeric -40 to 40 V

Table of Contents

Other manuals for Agilent Technologies 4155C

Related product manuals